StatSpin® Express 4 Horizontal Centrifuge Operator’s Manual A Division of IRIS International, Inc.
6 Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover
7 2. Locate 1/8 inch hex key supplied and remove center screw of rotor hub. 3. Remove rotor lifting straight up and out of centrifuge. To re-ins
8 Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenanc
9 Periodic Maintenance Check When the cycle counter in the instrument reaches 50,000 spins, the Error/Service indicator will continuously illuminate
10 Troubleshooting Symptom: Solution: Red Error / Service indicator flashing, unit will not spin. Refer to Error Code Table of manual to identify a
11 Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. Instrument fuse(s) are located on the rear panel. Remove
12 Section 5 Appendix Specifications Product Number SSH4 Model Number M510 Spin Parameters 3 min: 180 seconds @ 5100 rpm / 4000 xg 5 min: 300 sec
13 References 1. CLSI. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Third Edition.” CLSI documen
14 Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the centrifuge, with the exception
Operator’s Manual StatSpin® Express 4 Horizontal Centrifuge Model Number M510 For in-vitro diagnostic use Product Number SSH4 – StatSpin® Exp
i Table of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installatio
ii How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the inf
1 Section 1 Unpacking and Installation Inspect Packaging The StatSpin® Express 4 and its accessories are delivered in a single carton. If the centr
2 Section 2 System Overview Principle and Intended Use For in-vitro diagnostic use to produce the rapid separation of whole blood contained in origin
3 Locked Locked position for shield Unlocked Unlocked position for shield Operator Controls Start The Start button initiates the selected cycle
4 Error Codes Table Failure to reach speed in 30 seconds. Confirm no obstruction in bowl area, contact customer service. Latch is not lock